TBT-2430A Automatic Freezing Point Tester

Transport Package: Standard Export Packing
Specification: ISO
Trademark: TBT
Origin: Nanjing, China
Customization:
Manufacturer/Factory, Trading Company
Gold Member Since 2010

Suppliers with verified business licenses

Jiangsu, China
Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
Years of Export Experience
The export experience of the supplier is more than 10 years
Experienced Team
The supplier has 8 foreign trading staff(s) and 3 staff(s) with over 6 years of overseas trading experience
to see all verified strength labels (20)

Basic Info.

Model NO.
TBT-2430A
HS Code
90158790
Production Capacity
500sets/Month

Product Description

TBT-2430A Automatic Freezing Point Tester
TBT-2430A Automatic Freezing Point Tester

Summary
The instrument is designed and made as per the Industrial Standard of People's Republic of China SH/T0090-1991(2000) Standard Test Method for Freezing Point of Engine Coolant. It can be used to test freezing point of engine coolants and condensation liquids.
 
I.Main technical features
  1. It adopts dewar and can realize the test process of sample observation.It is clear to read.
  2. It adopts the refrigeration compressor from Germany. The temperature of cold bath can reach to -70ºC with the cascade technology.
  3. PID temperature control technology, stable temperature control and the accuracy is ±0.1ºC.
  4. Advanced electric stirring .No noise and the fault rate is low.
  5. It adopts single chip controlling technology.It can finish all of the test and detection process automatically.
  6. It adopts touch screen,detection system with high accuracy and good repeatability. Easy to operate.
The instrument's best feature: compact structure,double test-position design. The whole process can test automatically,touch screen and easy to operate,high accuracy and good repeatability.
 
II. Main technical specifications
1.Power supply: AC 220 V±10%, 50 Hz;
2.Working bath: double dewar.
3.Freezing point range: -54ºC~2ºC. 
4.Cold bath measurement temperature: -100ºC~50ºC
5.Sample measurement temperature: -100ºC~50ºC
6.Temperature control accuracy: ±0.1ºC
7.Refrigerator system: imported cascade refrigeration compressor 
8.The refrigeration power: 800W
9.The heating power: 400W
10.Condensation liquids stirring: electric stirring.the power is 10W,1200RPM
11.Sample stirring: mechanical stirring, 60RPM.
12. Ambient temperature:(25~35)ºC.
13. Relative humidity:  ≤80%
14. Maximum power consumption: 2000 W
15.Dimension: 760mm×460mm×630mm

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now